To optimize the life of your handpieces, it is essential to lubricate them meticulously.
Indeed, if the manufacturer’s lubrication recommendations were not respected and there is breakage, it will not be covered under warranty.
Have a look at our handpiece lubrication tips to avoid any breakage.
1. Drops of oil here – daily.
2. Drops of oil here – before and after autoclaving cycle (or according to manufacture). Lubricate the high-speed handpieces daily on the smaller of the two large holes.
3. Oil after every 15 minutes of use.
4. Do not leave levers open.
5. Daily lubricate chuck.
1. Drops of oil at these two places – each use.
2. The low-speed handpiece can be broken down into three parts. Disassemble the three parts;
3. Lubricate both ends of the contra-angle.
4. Lubricate the prophy angle.
5. Re-assemble the handpiece.
For drop oil, with the bur inserted (for the high-speed), place lubricant (2-3 drops) into the air intake, run and purge unit for 20 seconds, or until all excess oil is expelled.
For aerosol oil, with the bur inserted (for the high-speed), spray lubricant for 1 second into the air intake and run on purge unit for 20 seconds, or until all excess oil is expelled.
If expelled oil is colored, repeat the lubrification process until expelled oil is clear.
+ Remove the burs and other attachments (never sterilize with attachment).
+ Lever-operated chucks must be closed.
+ Never cool a hot handpiece under cold water or by placing it in a refrigerator.
+ Allow drying cycle to complete before removing handpiece from the autoclave to ensure that all moisture has been removed.
+ Make sure the handpiece is completely cool and dry before operating.
+ Have autoclave unit serviced regularly by manufacturer’s representative.
+ Do not cheat the bur this will damage the chuck.
+ Do not use pliers to remove burs.
+ Use a proper oil adaptor for your aerosol can.
+ Set air pressure according to manufacture (usually between 30-40 psi).
HIGHDENT Quattro Plus Veterinary Dental Unit
tvet 11E
Fully automatic autoclave
Dental Table
Vet-Tables Scissor
These days, I focus most of my professional time teaching/mentoring doctors and technicians who want to become efficient and skilled (safe) anesthetists. I have tried
Table of Contents The Occupational Safety and Health Administration, commonly known as OSHA, is an agency that falls under the United States Department of Labor.
Opening a veterinary clinic certainly presents several challenges. Have you ever had doubts about the success of your dream project? Have you ever had doubts
You’ve just had the frantic period, you’ve experienced staff shortages, impatience from some of your customers, equipment breakdowns, etc. In other words, the number of
Previous Next A new configuration from Dispomed We are proud to present a new configuration of surgery top now available for our Vet-Tables! Our V-top
**All accessories may be cleaned with diluted alcohol or a mild cleaner on gauze or cotton** LW8 LITE and 9x Models ECG (Keep attached to